User manual
22 Cubis MSU Brief Instructions
Information for the User
y The procedure is executed; progress is shown on the display.
t Wait until the calibration/adjustment has been completed.
y When the procedure has ended, the selection list is shown again.
t Select Back to return the balance to the operating mode.
Calibration/Adjustment
with External Calibration Weight
h
This function requires an external calibration weight. Please note the
tolerance of the calibration weight you use.
t Switch to the operating mode, if necessary.
t Make certain that the weighing pan is empty.
t Select Cal./Adj.
t Select External cal./adj., factory-defined weight followed by Start
t As soon as you are prompted to do so, lay the calibration weight on the
balance.
y The procedure is executed; progress is shown on the display.
t Wait until the calibration/adjustment has been completed (approximately
15 seconds).
y When the procedure has ended, the selection list is shown again.
t Select
Back to return the balance to the operating mode.