User manual

Information for the User
Cubis MSU Brief Instructions 19
t Select the method you wish to use in your task: Buoyancy or Displacement.
t Select the density determination kit you will be using.
YDK density kit from Sartorius: For analysis with the YDK kit.
Other density kit: If you would like to use a different kit
No density kit, no correction: If you are not using a kit.
t Enter values as prompted (see below for details).
Input Prompts/Settings for the YDK01MS Sartorius Density Determination Kit
and for Density Determination Without a Kit
*= Factory Settings
Liquid
*Water: For determining density relative to water
Ethanol: For determining density relative to ethanol
User input: For density determination relative to a different liquid; enter both the liquid and its
density at the indicated temperature.
Density at 20 °C/68 °F and exp. coefficient: For density determination relative to a different liquid; enter the liquid, its density
at 20 °C/68 °F and the expansion coefficient. The density at the indicated
temperature is calculated automatically.
Air Density
*Air density at 20 °C/68 °F: For analysis under standard laboratory conditions
User input: For analysis under other conditions; enter the air density.
Number of Decimal Places
No decimal places: The analysis result is displayed without decimal places.
1, *2, 3 or 4 places: The analysis result is displayed with the selected number of decimal places.
Print Output
No printout The results of the measurement are not printed.
*All results and parameters The results of the measurement and all configured parameters are printed.
Input Prompts/Settings for Other Density Determination Kit
Parameters for the Density Kit
Name: Enter the name of the kit
Number of wires: Enter the number of wires in the density kit
Wire diameter: Enter the diameter of the wires in mm
Vessel diameter: Enter the diameter of the vessel in mm
All other prompts are the same as those for the Sartorius density kit (see above).