Operating instructions

Configuring the Scale
Purpose
You can configure your FCT01-X
terminal to meet individual
requirements by entering user data
and setting parameters in the Setup
program.
The Setup menu contains the
following submenus:
Balance/scale functions
Device parameters
Application parameters
Printout
Device information
Language
Factory settings
Configuring the Scale for Use in
Legal Metrology
Set the menu access switch as
described below to configure the
following functions for use of the scale
in legal metrology:
Display: Verification scale interval
e; lower limit of the weighing
capacity
Min
External calibration blocked
Preparation
§ Remove the cap from the back of
the scale housing
§ Move the switch to the left.
> When the switch is in the left-hand
position, the Setup menu is locked
and the scale can be used in legal
metrology
> When the switch is in the right-
hand position, the menu is
accessible
Setting the Language
You can choose from 5 languages for the information display:
German
English (factory setting)
English with U.S. date/time format
French
Italian
Spanish
Example: Settting the Language to “U.S. Mode”
Step Press key Display/Output
(or follow
instructions)
1. Select Setup menu M
2. Select Language repeatedly
and confirm press
q soft key,
then
O soft key
3. Select U.S. mode Q soft key
4. Save language l soft key
5. Exit the Setup menu oo soft key
7
Configuring the Scale