Operating instructions
Configuring the Scale
Purpose
You can configure your FCT01-X
terminal to meet individual
requirements by entering user data
and setting parameters in the Setup
program.
The Setup menu contains the
following submenus:
– Balance/scale functions
– Device parameters
– Application parameters
– Printout
– Device information
– Language
– Factory settings
Configuring the Scale for Use in
Legal Metrology
Set the menu access switch as
described below to configure the
following functions for use of the scale
in legal metrology:
– Display: Verification scale interval
e; lower limit of the weighing
capacity
Min
– External calibration blocked
Preparation
§ Remove the cap from the back of
the scale housing
§ Move the switch to the left.
> When the switch is in the left-hand
position, the Setup menu is locked
and the scale can be used in legal
metrology
> When the switch is in the right-
hand position, the menu is
accessible
Setting the Language
You can choose from 5 languages for the information display:
– German
– English (factory setting)
– English with U.S. date/time format
– French
– Italian
– Spanish
Example: Settting the Language to “U.S. Mode”
Step Press key Display/Output
(or follow
instructions)
1. Select “Setup” menu M
2. Select “Language” repeatedly
and confirm press
q soft key,
then
O soft key
3. Select “U.S. mode” Q soft key
4. Save language l soft key
5. Exit the Setup menu oo soft key
7
Configuring the Scale