Operating instructions
63
Operating the Scale
Practical Example
When weighing in formulation components, the second component is over-poured.
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Recalculation.
Setup: Printout: Autoprint upon initialization: Only main values
Step Press key(s) Display/Output
(or follow instructions)
1. Delete previously stored values, c
if necessary
2. Place container for filling Place empty container
components on the scale on the scale
3. Tare the scale t
4. Add the first component Weigh the first component
into the container
5. Store component Press ther
Comp.1 soft key
6. Add the second component Weigh the second
component into the container
7. Start recalculation, because
Recalc soft key
10.73 g were not poured,
rather only 10.60 g
8. Either press the minus or plus
Minus soft key repeatedly
key to correct the value…
Comp1 + 25.08 g
s2
s2
s2
s2