Operating instructions

62
Factory Settings
Printout: Application-defined output:
Auto print upon initialization:
All values
Line format:
for other apps/GLP
(22 characters)
Soft Key Functions
Comp.xx Store component
weighed-in
Add.xx Store weighed-in
component in additive
weighing mode
Div. Store divisor before or
during component
weighing
Recal. Start correction
procedure for
recalculation
UAdd./ Toggle display between
UComp. component weight and
total weight (additive
mode)
Comp. Store numeric input for
recalculation
Minus Set value given for the
formula
Plus Set value given for the
formula
Preparation
Turn on the scale: press e
> The Sartorius logo is displayed
Select the “Recalculation” application in the Setup menu: press M
Select
Application parameters: press the q soft key 2 x, then
the O soft key once
Select Application 1 (basic settings): press the O soft key
Select Recalculation: press the Q or q soft key, repeatedly if necessary
Confirm Recalculation: press the o soft key
Save settings and exit the Setup menu: press the oo soft key
Additional Functions
In addition to the functions for:
alphanumeric input,
taring (not during alphanumeric
input), and – if recalculation has
been completed (otherwise Err 10
will be shown) –
printing,
you can also access the following
functions from this application:
Calibration/Adjustment
Press the
isoTST soft key
> See “Calibration/Adjustment”
for further instructions
Toggling to the Next Application
Press D
> See the section on the
corresponding application
program for further instructions
Setup (setting parameters)
Press M
> See “Configuring the Scale” for
further instructions
Turning Off the Scale
Press e
> The scale shuts off
> The display goes blank, then OFF