Operating instructions

61
Operating the Scale
Recalculation
Purpose
With this application program you
can compensate for over-poured
components in formulation.
If a component is over-poured when
weighing in the individual formulation
components, the mixture already
poured cannot be used in its current
composition. To avoid having to
discard the materials weighed, you
can adjust the proportions of the
formulation to compensate for the
over-pour.
When you use this application, the
scale mainly controls the recalculation
procedure.
You can use this application program
in combination with a program
chosen from Application 2
(checkweighing, time-controlled
functions) and one from Application 3
(totalizing, statistics) as well as with
the other extra functions.
Features
Individual components (up to 99)
weighed in with a readout
showing from »0« to the desired
component weight
Transaction counter shows the next
component expected
Weighed components are stored,
followed by automatic printout and
taring
Additive weighing of components
with printout
Toggle the display between
component weight and total
formulation weight (additive mode)
after first component is stored
Stored component weight displayed
as true net weight for 2 seconds
Enter a divisor before or during
component weighing. For
example, if the formulation has a
total weight of 100 g, enter the
divisor 10 to weigh in a total
formulation of 1,000 g
If a component is over-poured, you
can use the recalculation function
to change the amount of this
component indicated in the
formulation by using plus or minus
keys or numeric input; the scale
calculates a factor by which all
components amounts will then be
adjusted
Recalculation factor displayed
in the text line, with a warning
symbol if the factor is not equal
to 1
All components displayed with
number and the amount (by
weight) to be added in follow-on
filling; components displayed in
sequence by the scale.
Display of actual net weight during
follow-on filling
After the amounts of the components
already weighed have been
corrected, weighing continues
according to the adjusted
formulation amount; the readout is
recalculated (updated) according
to the divisor
You can repeat the over-pour
correction procedure as often as
necessary, in case other
components are over-poured
After follow-on (corrective) filling,
the total amount differs from that
given for the formulation, but the
proportion of components in
relation to each other is the same
You can have the weight printed
after each measurement
Individual component weights are
printed as “Compxx.”
Press c to exit the application
program. The component memory
is cleared and the sum of compo-
nents printed as “S-Comp.”
Toggle between the recalculation
program and other applications
(e.g., checkweighing) by
pressing D.