Operating instructions
52
Practical Example
Determining an Unknown Piece Count; Weighing-In the Preset Reference Sample Quantity
Settings (changes in the factory settings required for this example):
Setup: Application parameters: Application 1: Counting: Average piece weight updating: Manual
Setup: Application-defined output: Auto print upon initialization: All values
Step Press key(s) Display/Output
(or follow instructions)
1. Delete previous values if necessary c
2. Prepare a container for the Place the empty container
parts to be counted on the scale
3. Tare the scale t
4. Place the reference sample quantity Place the displayed number
on the scale of parts in the container
(example: nRef = 10 pcs)
5. Determine the average piece
Start soft key
weight (Number of decimal
places displayed depends on
the scale model)
6. If necessary, increase the number Place additional parts
of parts and perform reference in the container
sample updating (example Update. soft key
7 additional parts)
7. Weigh uncounted parts Place parts to be counted
in the container
8. If desired, print total piece count p
(here: 153 pcs)
nRef + 10 pcs
wRef + 2.14800 g
Qnt + 153 pcs
nRef + 17 pcs
wRef + 2.14800 g