Operating instructions
46
Calibration/Adjustment Printout
Data Block Printout
You can have the results of a
calibration/adjustment procedure
printed out. You can configure
whether the printout is generated as
soon as the procedure is completed,
or whether a number of calibration/
adjustment procedures (up to 50) are
collected for a data block printout.
Data Block Printout of Calibration/
Adjustment Data
With the following Setup menu confi-
guration, you can store the data from
up to 50 calibration/adjustment pro-
cedures and have it printed on
request:
– GLP/GMP adjustment record
On request, from
record memory
When the memory contains 50 data
records:
– additional records are output
automatically
If at least one data block printout data
record has been configured, the
following soft keys are available after
you press the CAL/isoTST soft key:
Info The number of records is
displayed in the text line
PrtPro Print accumulated records
DelPro Delete accumulated
records; records can only
be deleted after a printout
has been generated. If a
password has been
assigned in the Setup:
Input menu, you must enter
either the configured
password or the General
Password (40414243)
before you can delete the
records.
For internal calibration/adjustment,
the initialization mode of the
procedure is printed in the
Start
line.
---------------------
13.05.2000 09:17
SARTORIUS
Model FCT01-X
Ser. no. 60419914
Vers. no. 01-41-02
ID
--------------------
24.04.2000 12:03
Start: manual
Diff. + 0.01 g
External calibration
completed
27.04.2000 11:53
Start: Ext.cal.
WID
Nom + 200.00 g
Diff. + 0.01 g
External adjustment
completed
Diff. + 0.00 g
---------------------
13.05.2000 09:17
Name:
---------------------
GLP header
List of Calibration/Adjustment
Procedures:
Example 1:
External calibration
Example2:
External calibration/adjustment
GLP footer