Operating instructions
44
Calibration and Adjustment
Sequence
In the Setup menu, you can configure
the scale so that:
– calibration is always followed
automatically by adjustment
Calibrate, then auto
adjust or
– you have the choice of ending the
sequence or starting adjustment
after calibration
Calibrate,
then manual adjust
(factory setting)
If no deviation is determined in
calibration, or the deviation is within
the tolerance limits dictated by the
degree of accuracy you require, it is
not necessary to adjust the scale. In
this case, you can end the calibration/
adjustment sequence after calibration.
There are 2 soft keys active at this
point:
–
Start to start adjustment
– End to end the sequence
External Calibration/Adjustment with
a User-Defined Calibration Weight
In the Set up menu (Balance/scale
functions: Calibration/adjustment:
CAL/isoTST key function), you need
to set
Ext. cal./adj.;
user-defined wt.
or select
this using the
Selection mode
(factory setting).
You can define a weight for calibrati-
on/adjustment. External calibrati-
on/adjustment must be performed
with weights that are traceable to a
national standard and that have error
limits that are no greater than 1/3 of
the required tolerance of the display
accuracy.
Sequence of external calibration
(adjustment: see left column). First
select External adjustment:
User-def. wt.
The scale has a factory-set weight
value (see “Specifications”).
To reset a user-defined calibration
weight to the original factory setting:
$ Enter the factory-defined value
manually (see next page).