Operating instructions

Generate “FlexPrint”
Purpose
You can use a PC with the YAD02IS
"Nice Label Express" software from
Sartorius to load user-defined label-
printing formats in your FCT01 scale.
This software lets you connect any of
a number of printers, equipped with a
variety of printer fonts, to the RS-232
interface on your FCT01 terminal.
Features
With the FlexPrint option activated:
Print command generates configured
printout (if print instruction file exists;
see table, next page).
Print command generates default
printout (if print instruction file does
not exist).
The function that generates an
automatic printout upon initialization
of an application cannot be used.
Initialization data can be output
only to a print instruction file.
The following items are output only as
standard printouts:
Calibration/adjustment reports
SETUP printouts
p key in Identifier mode:
display page
p key in Product data memory
mode: display page
Parameter Display text Indiv. Comp. Total
With the “Differential
weighing”application:
Lot name
Lot name xx
Sample number Sample no. x
Date/time of
sampling Sample date x
Sample
identification Sample ID x
Tare weight or input Tara (T/PT) x
Initial weight
or input Net initial wt. x
Backweight
or input;
residue as weight Backweighed res. x
Residual weight
in percent Residue in % x
Weight loss Weight loss x
Loss in percent Loss in % x
Factor used in
calculation Factor x
Parameter Display text Indiv. Comp. Total
Loss calculated
D.res. x
as a weight
Ratio1 % (DR) Ratio1 x
Ratio2 % (OR) Ratio2 x
Date/time of
statistics Date of statis. x
Statistics
identification Statistics ID x
Number of
samples No. of samples x
Mean value Mean value x
Standard
deviation Standard dev. x
Variation
coefficient Variat. coeff. x
Sum Sum x
Minimum Minimum x
Maximum Maximum x
Difference between
minimum &
maximum Difference x
30
p key in Parameter display
page mode for Checkweighing
application: print function carried
out
MR function carried out when p
key pressed during evaluation (info
window) in Totalizing and Statistics
applications.
Printout for legal metrology:
Weight blocks (special weight
value formats that are acceptable
in legal metrology) are designated
by the following line, which is
printed in both header and footer
and cannot be edited:
------ CE [M] -------
Examples of Weight Block Printouts
Without tare:
------ CE [M] -------
N + 348.65 kg
------ CE [M] -------
With tare:
------ CE [M] -------
G + 459.70 kg
N + 348.65 kg
T1 + 111.05 kg
------ CE [M] -------
Scale tare (2nd tare memory):
------ CE [M] -------
G + 124.45 kg
N + 100.00 kg
T1 + 24.00 kg
T2 + 0.45 kg
------ CE [M] -------
Preparation
Configure FlexPrint in Setup:
press the M key
Select Printout:
press q soft key 3 times and then
O soft key
Confirm
Application-
defined output
:
press O soft key
Select FlexPrint:
press
q soft key 4 times and then
O soft key
Select On:
press q soft key and then
O soft key
Save settings and exit Setup:
press