Error Codes
Error Code/Message Displayed Cause Solution
Cannot update Reference updating not possible See “Counting” in “Operating
(with the counting application) the Scale” for reference
updating criteria
Not a number Input wrong (with any Follow the instructions for the
xxxxx Too low application program), e.g., application programs
xxxxx Too high alphabetic input not allowed
Too many char. Input text too long Allowable text lengths, incl.
decimal point:
– S ID, NUM, L ID,
ID: max. 20 characters
– W ID: max. 14 characters
Wrong line format Configured printout, printout memory Select the 22-character format:
and 16-character format selected
Limits unequal Unit entered for tolerance limits Adjust tolerance
for unit in checkweighing different limits
from the application used
Equation too long Equation exceeds 28 characters Limit equation to 28 characters
Differential Weighing:
SAMPLE: Confirm delete/ “SAMPLE: delete/omit” prompt Select Yes to delete
omit on display page for samples Select Omit to omit
SAMPLE: Include “SAMPLE: include” prompt Select Omit to include a
on display page for samples sample already omitted
Cannot store File manager: Delete lot(s)
– Not possible to save data
– No available memory
Cannot load File manager: Delete lot(s)
– Not possible to load data
– Memory capacity limit reached
Only 30 backweights An attempt was made to save a None
possible 31st backweighing operation
LOT: already exists Lot already exists on the display Choose a different lot ID
page for LOTS
No sample
The Sample key was pressed when Save sample first
the display page for Lots was shown,
but there are no samples in the
lot selected
Out of range On the display page for LOTS Enter the correct lot or
or SAMPLES, an alphanumeric sample ID
lot or samle ID was input and
not found
Not enough memory space An attempt was made to save Useless memory or
or more than 999 samples using delete one or more lots
999 samples maximum the numeric keys and the #Spl
key
Sample omitted An attempt was made to save None
data from an omitted sample
Value too small An attempt was made to Place the particular weight
to accept save a tare, initial or backweight on the scale
that is less than one display digit
No choice available Factor was selected while Selection not possible
attempting to activate the
display page for results