Operating instructions
148
Set/Delete Preload
The preload can only be set or deleted when the service switch is off (after opening
the switch, turn the display unit off and on again once). After you exit the adjustment,
adjust the service switch to “on” and activate the display unit by turning it off and back
on again once. This activates the L/H display.
The preload adjustment is dependent on the type of platform connected.
Set Preload
S-CAL: Set preload:
Press key Display Action
0.000 kg Place weight on pan: 8 kg
“Start” 8.000 kg
0.000 kg Press CF to end routine
Clear Preload
S-CAL: Clear preload:
Press key Display Action
0.000 kg Remove weights from pan
– 8.000 kg
“Start” – 8.000 kg
0.000 kg Press CF to end routine
Determine User-defined Weight
S-CAL: Determining user-def. wt.:
Press key Display Action
0.000 kg
“Start” – 3.000 kg Place weight on pan:
e.g. 20 kg
+17.000 kg Wait until +17.000 kg is displayed
“Start” +17.000 kg
20.000 kg Press CF to end routine
Display *10
S-CAL: Display *10:
Press key Display Action
0.000 kg
“Start” 0.0000 kg
“End” 0.000 kg Routine ended