Operating instructions

147
Service
Enter Operator’s Weights
Weighing parameters (linearization weights can be changed only in the service mode)
Setup
Device parameter
ADC configuration (only visible in the service mode)
Calibration/adjustment
Function CAL key
Parameter for external weight (Example:)
Wt. ID (W ID): ABC001 Enter 14 characters max.
e.g. (Class M1)
Cal./adj. wt.: 25.000kg
Lin. wt. 1: 10.000kg
Lin. wt. 2: 15.000kg
Lin. wt. 3: 20.000kg
Lin. wt. 4: 30.000kg
S-CAL Functions (sequence of the lin./calibr. and adjustment routine)
Perform external linearization with user-defined weights
S-CAL: Ext. lineariz.; user-def. wt.
Press key Display Action
“TARE” 0.000 kg
“Start” 0.000 kg
–10.000 kg Place weight on pan: 10 kg
10.000 kg Wait until 0.00X kg is displayed
“Start” + 0.002 kg
–15.000 kg Place weight on pan: 15 kg
15.000 kg Wait until 0.00X kg is displayed
“Start” +0.004 kg
20.000 kg Place weight on pan: 20 kg
20.000 kg Wait until 0.00X kg is displayed
“Start” + 0.005 kg
30.000 kg Place weight on pan: 30 kg
30.000 kg Wait until 0.00X kg is displayed
“Start” 0.009 kg
+ 0.000 kg Remove weights from pan
0.000 kg Wait until the adjustment routine
0.000 kg has been completed
S