Operating instructions

144
Example 1: Configuring the A/D Converter with a Weighing Platform Connected (such as a Sartorius IWA platform)
Step Press key(s) Display/Output
(or follow instructions)
1. Remove the cap from the back
of the scale housing and move
the switch to the right
2. Turn the terminal off and back on
3. Enter the service password and * * * * * * * *M
select Balance/scale functions
4. Select ADC configuration O soft key twice
5. Select Standard and confirm O soft key
6. Select Ranges and confirm O soft key
7. Select Single-range O soft key
scale and confirm
8. Enter and confirm the scale 0 . 1 and l soft key,
interval d and maximum capacity
(in this example, 0.1 and 300 kg) 3 0 0 and l soft key
9. Save the data entered , soft key and
then
Yes soft key
10. Calibrate/adjust the scale: see
Operating the Scale in this manual,
or refer to the service manual
11. Return the menu access switch to the
left-hand position
12. Turn the terminal off and back on e key