Operating instructions
140
Verification of Intrinsic Safety
8
Blatt
Page
8
von
of
Maßstab /
Scale
---
FCT01-X.. /
Ausgabe / Revision
01
Freigabe
Released by
Geprüft
Reviewed by
Erstellt
Written by
25.08.05
25.08.05
25.08.05
Datum
Date
Klausgrete
Klausgrete
Klausgrete
Name
Material
65638-099-60-A4
Benennung / Title
Änderung / Alteration
---
Zeichnungs-Nr. / Drawing number
FCT01-X..
II 2 G EEx ib IIC T4
Control Lines, Separate Foot Switch
Non-hazardous
area
Hazardous area
Zones: 1, 2
Gas: IIC T4
RS-232
Data interface
port
Notes:
1: Per circuit
2: Combined circuits
*: Option; only passive wiring (not zone 20), can also be connect ed by a separate cable gland
YPE05-X
*
Foot switch
Barrier
YDI03-Z
Circuit: RS-232
1)
Control signals
1)
Vo 12 V 12 V
Io 82 mA 7.5 mA
Po 240 mW 22 mW
Co 1.41 µF 1.41 µF
Lo 5.5 mH 5.5 mH
Lo/Ro 144 µH/ohm 1.5 mH/ohm
Data interface RS-232 Control signals
port
(pins A/J/K/N and M)
1)
(pins C/D/E/F/G and M)
1)
Vi 12.6 V 12.6 V
Ii 85 mA 10 mA
Pi 270 mW 30 mW
Ci 3 nF 4 nF
Li 0 0
Cable: install as stationary
and protected equipment!
Maximum cable length 70 m
for a 10-wire standard cable
with a maximum of 200 nF/
km and 1 mH/km.
The length of the installed
cable is defined as 20 m
because of the RS-232
specifications.
System assessment
(see EC-Type Examination Certificate KEMA
02ATEX1108 X):
Co <= 125 nF; Lo <= 100 µH; Lo/Ro <= 33 µH/W