Data Input Format
You can connect a computer to your scale to send commands via the scale
interface port to control scale functions and applications.
The commands sent are control commands and may have different formats;
e.g., control commands can have up to 26 characters. Each character must
be transmitted according to the settings configured in the Setup menu for data
transmission.
Formate for Control Commands
Format 1: Esc ! CR LF
Format 2: Esc ! # _ CR LF
Format 3: Esc ! # & (max. 20 &) & _ CR LF
Format 4: Esc ! # & (max. 8 &) & _ CR LF
Esc: Escape _: Underline (ASCII: 95)
!: Command character CR: Carriage Return (optional)
#: Number LF: Line Feed (optional)
&: Number or letter max: depends on command
character; i.e., parameter:
ASCII code chart once the max. length is
see chapter on “Overview” reached, input received is
cut off, rather than rejected
as with keyboard input
Format 1
! Meaning
K Weighing mode 1
L Weighing mode 2
M Weighing mode 3
N Weighing mode 4
O Block keys
P Print
R Unblock keys
S Restart
T Tare and zero
Z Internal calibration/adjustment
Format 2
!# Meaning
f3 Zero
f4 Tare (without zeroing)
kF1 Soft key 1* Function depends
... on setting in appli-
kF6 Soft key 6* cation program
kF7 M function key
kF8 D function key
s3 c function key
x0 Perform internal calibration
x1 Print scale model
x2 Print serial no. of weighing
x3 Software version of weighing
platform
x4 Software version of display &
control unit
x5 Print (GLP) scale ID no.
x6 Print weight set “inventory” no.
x7 Print weighing series no.
Format 3 (not allowed in the
Setup menu)
!# Meaning
z5 Input (GLP) scale ID no.
z6 Input weight set “inventory” no.
z7 Print weighing series no.
Format 4
! Meaning
t Text input in display
* numbered from right to left