Operating instructions
125
Operating the Scale
ISO/GLP Printout
You can have the parameters pertai-
ning to weighing conditions printed
before (GLP header) and after (GLP
footer) the values from the weighing
series (Setup: Printout: ISO/GLP/
GMP printout: Always on). These
parameters include:
GLP header:
– Date
– Time at the beginning of a weig-
hing series
– Scale manufacturer
– Scale model
– Model serial number
– Software version
– ID number
GLP footer:
– Date
– Time at the end of the weighing
series
– Field for operator signature
Operating the Scale with an
ISO/GLP-compliant Logging Device
(Printer)
ISO/GLP-compliant documentation
requires a computer with special
software. Contact Sartorius for
a detailed description.
Setting:
Setup: Printout: ISO/GLP/GMP
printout: Always on
The record is output to a Sartorius
YDP03-0CE Data Printer or a
computer.
End GLP printout:
● Press c
End GLP printout while application
is active:
This requires the following settings:
Setup: Device settings: Keys: CF
function in application: Clear only
selected applications
● Press c
> Text line: CF selected:
clear application
● Press the GLP soft key
--------------------
17.01.2000 16:12
SARTORIUS
Model FCT01-X
Ser. no. 91205355
Vers. no. 01-42-02
ID 12345678901234
--------------------
L ID 12345678901234
nRef 10 pcs
wRef 1.35 g
Qnt + 235 pcs
Qnt + 4721 pcs
S ID 12345678901234
Qnt + 567 pcs
--------------------
17.01.2000 16:13
Name:
--------------------
--------------------
17.01.2000 16:24
SARTORIUS
Mod. FCT01-X
Ser. no. 91205355
Vers. no. 01-42-02
ID
--------------------
Start: ext.cal.
WID
Nom + 200.00 g
Diff. + 0.06 g
External adjustment
completed
Diff. + 0.00 g
--------------------
17.01.2000 16:25
Name:
--------------------
Dotted line
Date/time
Scale manufacturer
Scale model
Scale serial number
Software version (display and control unit)
Scale ID no.
Dotted line
Weighing series no. (lot)
Application initialization value
Application initialization value
Counting result
Counting result
ID for counting result
Counting result
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line
Record of Internal
Calibration/Adjustment:
Dotted line
Date/time
Scale manufacturer
Scale model
Scale serial number
Software version (display and control unit)
Scale ID no.
Dotted line
Start mode for ext.calibration/adjustment
Weight set number
Exact calibration weight value
Difference after calibration/adjustment
Confirmation of completed calibration/
adjustment routine
Difference between current and target
values after calibration
Dotted line
Date/time
Field for operator signature
Blank line
Dotted line