Operating instructions

116
Data Output Functions
There are 3 options for data output:
Output to the display and control unit
Output to a printer (generate a printout)
Output to a peripheral device (e.g., computer) via the interface port
Output to the Display and Control Unit
The display is divided into 9 sections. Information about the scale,
the application being used and the sample weighed is output in the
following sections:
Line for metrological data
Bar graph
Plus/minus sign, stability symbol display
Line for measured values (weights)
Weight unit display
Data in tare memory; calculated value
Application symbol display
Text line
Soft key labels
Line for Metrological Data (on scales verified for legal metrology)
This line shows:
Maximum weighing capacity (such as 6200 g)
Minimum weighing capacity; the weight must not go below this limit
(such as 5 g)
Verification scale interval; irrelevant if the scale is not used in legal metrology
(such as 0.1 g)
Readability/actual scale interval: indicates the scales display increment in
digits (such as 0.01 g)
Bar Graph (overview display)
In the bar graph, weighing results are displayed either
as a percentage of the maximum scale capacity, or
in relation to a target value, with tolerance limits indicated.
You can turn off (blank) the bar graph display (Setup: Device parameters:
Display: Digit size: 13 mm + text display or 13 mm)
Plus/Minus Sign, Stability Symbol
This section shows:
–“Busy symbol
Plus or minus sign
Zero symbol (indicating the scale has been zeroed)
Max6200 g
Min 5 g
e= 0.1g
d= 0.01g
J
F H
S