Operating instructions

108
Saving Values Manually in M+
Purpose
This extra function Manual storage in
app.3 memory enables you to load
weight values and calculated results
directly from Application 1 (such as
counting, weighing in percent) or
Application 2 (checkweighing, time-
controlled functions) into Application
3 (totalizing, formulation, statistics).
Features
You can assign this function to the
fourth or fifth soft key (from the
right); i.e, F4 or F5; the soft key
label for this function is M+
An Application 3 program
(totalizing, formulation, statistics)
must be running so you can
display and print the result
Factory Settings
No user-definable parameters
Preparation
Turn on the scale: press e
> The Sartorius logo is displayed
Select Extra function (F4) or Extra function (F5) in the Setup menu:
press M
Select
Application parameters: press the q soft key 2 x,
then the O soft key
Select Extra function (F4) or Extra function (F5):
press the q soft key 3 x (or 4 x), then the O soft key once
Select Man. store in app.3 memory (M+)
Confirm Man. store in app.3 memory (M+)
See also the Application Parameters (Overview) in the chapter entitled
Configuring the Scale
Save settings and exit the Setup menu: press the oo soft key