Operating instructions

Description of the Keys
e On/off key: switches the balance/scale
on and off or switches it to the
standby mode.
Battery operation: on; turns
backlight on; off
w Press less than 2 seconds:
zeros the balance/scale
Press and hold 2 seconds:
opens the application menu
q Starts calibration/adjustment
v Starts an application program
Scrolls in the application, configura-
tion and in the calibration menus
Confirms the selected setting;
exits application, configuration &
calibration menu if key is pressed
and held for at least 2 seconds.
r Generates a printout or data output
7
Operation
Basic Weighing Function
Features
Zeroing the balance/scale
You can zero the balance/scale within
the entire weighing range, up to the
maximum capacity.
Preparation
§ Switch on the balance/scale:
press the e key
§ If necessary, zero the balance/scale:
press the w key
$ If necessary, change the configuration
settings: see the chapter entitled
“Configuration”
$ If desired, load the factory settings:
see the chapter entitled
“Configuration,” menu item 9.– 1
Additional functions
$ Switching off the balance/scale:
press the e key
$ Battery operation: automatic shut-off
after 2, 5 or 10 minutes; see chapter on
configuration. Example: 2 minutes. If
the weight readout remains unchanged
and no keys are pressed for at least two
minutes, the battery symbol “
starts flashing. After another 2 seconds,
the balance/scale shuts off automati-
cally, unless a key is pressed.
Example: Determine weight of sample
Step Key (or instruction) Display
1. Switch on the balance/scale e
Self-test runs
Display: Software version
r 31.01 g
2. Place container on the balance/scale 52.0 g
(in this example, 52 g)
3. Zero the balance/scale w 0.0 g
4. Place sample in container on balance/scale 150.2 g
(in this example, 150.2 g).
Enter
CF