Operating instructions

Error Codes
16
Error codes are shown on the main display for approx. 2 seconds.
The program then returns automatically to the previous mode.
Display/Problem Cause Solution
No segments appear No power available Check the power supply
on the display
The AC adapter is not plugged in Plug in the AC adapter
Battery is drained Replace battery; recharge
battery using external charger
oL The load exceeds the Unload the balance/scale
balance/scale capacity
UL Weighing pan not in place Place the weighing pan
on the balance/scale
Something is touching Move the object that is
the weighing pan touching the weighing pan
disErr Display overflow: Reduce load on the
Value cannot be shown balance/scale
on the display
calErr Calibration parameter not met; Calibrate only when zero
e.g.: is displayed
Balance/scale not zeroed Press w to tare the
balance/scale
Balance/scale is loaded Unload the balance/scale
appErr Weight is too light or there is Increase the weight
no sample on the balance/scale on the balance/scale
with application in use
PrtErr Data interface for Contact the Sartorius
printing is blocked customer service center
Bal.Err Balance/scale loaded or defective Unload balance/scale before
when power was turned on switching on or contact
Sartorius customer service
Sys.Err Balance/scale defective Contact Sartorius customer
service
Max. weighing capacity The balance/scale was Place the weighing pan
is less than indicated switched on without on the balance/scale and
under “Specifications” the weighing pan in place press e
The weight readout The balance/scale was not Calibrate/adjust the
is obviously wrong calibrated/adjusted before weighing balance/scale
Balance/scale not zeroed Zero the balance/scale
If any other errors occur, contact your local Sartorius customer service center.