Operating instructions

Specific Gravity
Purpose
Use this application program to determine the specific gravity of a sample. The result is
displayed with one decimal place. Beaker and wire not included with balance/scale.
Example: Determine the specific gravity of a solid.
Step Key (or instruction) Display
1. Select application program w > 2 sec
2. Select Specific Gravity v repeatedly
3. Confirm setting
Symbol “ ” on the display:
application is active
4. If necessary, zero the balance/scale w
5. Start application program v
6. Confirm the display,
airval
7. Determine weight of the sample in air:
place sample on the balance/scale
8. Store weight-in-air value
9. Remove sample from balance/scale
10. Determine weight of sample in liquid:
connect wire and set up beaker
11. Confirm the display “Water”
12. Place sample in liquid
13. Store the weight-in-liquid value
and view the result
14. Clear the display w
15. Exit the Specific Gravity application > 2 sec
16. Reactivate Specific Gravity (if no other v
application program has been selected)
17. Repeat procedure starting from Step 5.
13
Enter
CF
Enter
CF
Enter
CF
Enter
CF
Enter
CF
Enter
CF