Operating instructions
Cubis MSE Operating Instructions 99
Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Modelle MSE12201S- MSE8201S- MSE5201S- MSE70200S- MSE36200S-
0CE 0CE 0CE 0CE 0CE
Accuracy class* K K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 100 100 100 1,000 1,000
Weighing capacity max* g 12,200 8,200 5,200 70,200 36,200
Calibration value e* g 1 1 1 10 1
Min. load min* g 5 5 5 50 50
Tare equalization range (subtractive) 100% vom maximalen Wägebereich
Application range according to DIR* g 5 – 12,200 5 – 8,200 5 – 5,200 50 – 70,200 50 – 36,200
Min. initial weight** g 100 100 100 1,000 1,000
Typical stabilization time s 0.8 0.8 0.8 1 1
Typical measurement time s 1 1 1 1.2 1.2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206 206 206 400 300 400 300 400 300
IP protection IP54
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26