Operating instructions
Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models MSE6202S- MSE6202P- MSE5202S- MSE4202S-
0CE 0CE 0CE 0CE
Accuracy class* K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* g 0.01 0.01/0.02/0.05 0.01 0.01
Weighing capacity max* g 6,200 1,500/3,000/6,200 5,200 4,200
Calibration value e* g 0.1 0.1 0.1 0.1
Min. load min* g 0.5 0.5 0.5 0.5
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.5 – 6,200 0.5 – 6,200 0.5 – 5,200 0.5 – 4,200
Min. initial weight** g 12 12 10 12
Typical stabilization time s 1 1 0.8 0.8
Typical measurement time s 1.5 1.5 1 1
Application range (temperature):
With “isoCAL” function +5 to +40 °C +5 to +40 °C +5 to +40 °C
Without “isoCAL” function +10 to +30 °C +10 to +30 °C +10 to +30 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206 206 206 140 140 206 206
IP protection IP54
Models MSE2202S- MSE1202S- MSE36201S- MSE36201P- MSE20201S-
0CE 0CE 0CE 0CE 0CE
Accuracy class* K K K K K
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 10 10 100 100/1,000 100
Weighing capacity max* g 2,200 1,200 36,200 10,200/36,200 20,200
Calibration value e* g 0.1 0.1 1 1 1
Min. load min* g 0.5 0.5 5 5 5
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.5 – 2,200 0.5 – 1,200 5 – 36,200 5 – 36,200 5 – 20,000
Min. initial weight** g 12 12 120 120 120
Typical stabilization time s 0.8 0.8 1.5 1.5 1.5
Typical measurement time s 1 1 2 2 2
Application range (temperature) With “isoCAL” function: +5 to +40 °C | Without “isoCAL” function: +10 to +30 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan dimensions (W + D) mm 206 206 206 206 400 300 400 300 400 300
IP protection IP54
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26
98 Cubis MSE Operating Instructions