Operating instructions

Technical Specifications
Model-specific Data
Verified models with EC Type Approval Certificate: Precision balances
Models MSE5203S-0CE MSE5203P-0CE MSE3203S-0CE MSE3203P-0CE
Accuracy class* k k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/2/5 1 1/10
Weighing capacity max* g 5,200 1,200/2,400/5,200 3,200 1,010/3,200
Calibration value e* mg 10 10 10 10
Min. load min* mg 100 100 100 100
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.1 – 5,200 0.1 – 5,200 0.1 – 3,200 0.1 – 3,200
Min. initial weight** g 1.5 1.5 1.5 1.5
Typical stabilization time s 1 1 1 1
Typical measurement time s 2 2 2 1.5
External standard calibration value g 5,000 (E2 5,000 (E2 2,000 (E2 2,000 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 ... +40 °C
Without “isoCAL” function: +15 ... +25 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
Models MSE2203S-0CE MSE2203P-0CE MSE1203S-0CE
Accuracy class* k k k
For verified models: EC Type-Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 1 1/10 1
Weighing capacity max* g 2,200 1,010/2,200 1,200
Calibration value e* mg 10 10 10
Min. load min* mg 100 100 100
Tare equalization range (subtractive) 100% from max. weighing capacity
Application range according to DIR* g 0.1 – 2,200 0.1 – 2,200 0.1 – 1,200
Min. initial weight** g 1.5 1.5 1.5
Typical stabilization time s 1 1 1
Typical measurement time s 1.5 1.5 1.5
External standard calibration value g 2,000 (E2) 1,000 (E2) 1,000 (E2)
(min. accuracy class)
Application range (temperature) With “isoCAL” function: +5 ... +40 °C
Without “isoCAL” function: +15 ... +25 °C
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height (draft shield DE)
mm 172
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26
96 Cubis MSE Operating Instructions