Operating instructions
94 Cubis MSE Operating Instructions
Technical Specifications
Model-specific Data
Verified Models with EC Type Approval Certificate: Micro- and ultramicrobalances
Model MSE6.6S-0CE MSE2.7S-0CE MSE3.6P-0CE
Accuracy class* k k k
For verified models: EC Type Approval Certificate D09-09-015, Type: MSX
Scale interval d* mg 0.001 0.0001 0.001/0.002/0.005
Weighing capacity max* g 6.1 2.1 1.1/2.1/3.1
Calibration value e* mg 1 1 1
Min. load min* mg 0.1 0.01 0.1
Tare equalization range (subtractive) 100 % from max. weighing capacity
Application range according to DIR* g 0.001 – 6.1 0.001 – 2.1 0.001 – 3.1
Min. initial weight** mg 2 1 4
Typical stabilization time s 5 7 5
Typical measurement time s 8 10 8
External standard calibration value
(min. accuracy class) g 5 2 3
Application range (temperature) With “isoCAL” function: +5…+40°C |
Without “isoCAL” function: +15 … +25°C
Display result (depending on the set filter level) By selection of 1 of 4 optimized filter levels
Weighing pan size mm 30 20 30
Weighing chamber height
(draft shield DM) mm 70 70 70
Protection Protected against dust and water
* DIR = Directive 90/384/EEC on non-automatic weighing instruments used within the European Economic Area
** = Typical min. initial weighing according to USP (Unites States Pharmacopeia), USP31-NF26