Operating instructions
Cubis MSE Operating Instructions 93
Technical Specifications
Model-specific Data
Precision balances
Models MSE20201S MSE12201S MSE8201S MSE5201S
Readability mg 100 100 100 100
Weighing capacity g 20,200 12,200 8,200 5,200
Tare range (subtractive) g – 20,200 – 12,200 – 8,200 – 5,200
Repeatability ±mg 70 50 50 50
Linearity ±mg 200 100 100 100
Corner load (test load [g]) mg 300 (5,000) 200 (5,000) 200 (5,000) 200 (2,000)
Min. initial weight* g 120 100 100 100
Sensitivity drift between +10 to +30°C ±ppm/K 2 4 4 4
Typical stabilization time s 1 0.8 0.8 0.8
Typical measurement time s 1.5 1 1 1
External standard calibration value kg 20 (F1) 10 (F1) 5 (F2) 5 (F2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206
Protection IP54
* = Position nach OIML R76
Models MSE70200S MSE36200S
Readability mg 1,000 1,000
Weighing capacity g 70,200 36,200
Tare range (subtractive) g – 70,200 – 36,200
Repeatability ±mg 500 500
Linearity ±mg 1,000 1,000
Corner load (test load [g]) mg 1,000 (20,000) 1,000 (10,000)
Min. initial weight* g 1,000 1,000
Sensitivity drift between +10 to +30°C ±ppm/K 3 2
Typical stabilization time s 0,8 0,8
Typical measurement time s 1 1
External standard calibration value kg 50 (F1) 20 (F1)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 400 300
IP protection IP54
* = Position according to OIML R76