Operating instructions

92 Cubis MSE Operating Instructions
Technical Specifications
Model-specific Data
Precision balances
Models MSE6202S MSE6202P MSE5202S MSE4202S
Readability mg 10 10/20/50 10 10
Weighing capacity g 6,200 1,500/3,000/ 5,200 4,200
6,200
Tare range (subtractive) g – 6,200 – 6,200 – 5,200 – 4,200
Repeatability ±mg 7 7/20/40 6 7
Linearity ±mg 20 50 10 20
Corner load (test load [g]) mg 20 (2,000) 50 (2,000) 10 (2,000) 30 (2,000)
Min. initial weight* g 12 12 10 12
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2 2
Typical stabilization time s 1 1 0.8 0.8
Typical measurement time s 1.5 1.5 1 1
External standard calibration value kg 5 (E2) 5 (F1) 5 2 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206 206 206 140 140 206 206
IP protection IP54
* = Position according to OIML R76
Models MSE2202S MSE1202S MSE70201S MSE36201S MSE36201P
Readability mg 10 10 100 100 100/1,000
Weighing capacity g 2,200 1,200 70,200 36,200 10,200/36,200
Tare range (subtractive) g – 2,200 – 1,200 – 70,200 – 36,200 – 36,200
Repeatability ±mg 7 7 70 70 70/500
Linearity ±mg 20 20 500 200 200
Corner load (test load [g]) mg 20 (1,000) 20 (500) 500 (20,000) 300 (10,000) 300 (10,000)
Min. initial weight* g 12 12 120 120 120
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 3 2 2
Typical stabilization time s 0.8 0.8 1 1 1
Typical measurement time s 1 1 1.5 1.5 1.5
External standard calibration value kg 2 (F1) 1 (F1) 50 (F1) 20 (F1) 20 (F1)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W + D) mm 206 + 206 206 + 206 400 300 400 300 400 300
IP protection IP54
* = Position according to OIML R76