Operating instructions

Cubis MSE Operating Instructions 91
Technical Specifications
Model-specific Data
Precision balances
Models MSE623S MSE623P MSE323S
Readability mg 1 1/2/5 1
Weighing capacity g 620 150/300/620 320
Tare range (subtractive) g – 620 – 620 – 320
Repeatability ±mg 0.7 1/2/4 0.7
Linearity ±mg 2 5 2
Corner load (test load [g]) mg 2 (200) 4 (200) 2 (200)
Min. initial weight* g 1.5 1.5 1.5
Sensitivity drift between +10 to +30°C ±ppm/K 2 2 2
Typical stabilization time s 0.8 0.8 0.8
Typical measurement time s 1 1 1
External standard calibration value g 500 (E2) 500 (F1) 200 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 140 140
Weighing chamber height
(draft shield DE) mm 172
Protection Protected against dust and water
* = Position according to OIML R76
Models MSE14202S MSE14202P MSE10202S MSE8202S
Readability mg 10 10/20/50 10 10
Weighing capacity g 14,200 3,500/7,000/14,200 10,200 8,200
Tare range (subtractive) g –14,200 –14,200 –10,200 –8,200
Repeatability ±mg 10 10/20/40 7 7
Linearity ±mg 30 50 20 20
Corner load (test load [g]) mg 20 (5,000) 40 (5,000) 20 (5,000) 20 (5,000)
Min. initial weight* g 15 15 12 12
Sensitivity drift between +10 to +30°C ±ppm/K 1.5 1.5 2 2
Typical stabilization time s 1 1 1 1
Typical measurement time s 1.5 1.5 1.5 1.5
External standard calibration value kg 10 (E2) 10 (E2) 10 (E2) 5 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4
Weighing pan size (W D) mm 206 206
IP protection IP54
* = Position according to OIML R76