Operating instructions
Technical Specifications
Model-specific Data
Microbalances 0.001mg
Model MSE6.6S MSE6.6S-F MSE3.6P
Readability mg 0.001 0.001 0.001/0.002/0.005
Weighing capacity g 6.1 6.1 1.1/2.1/3.1
Tare range (subtractive) g -6.1 -6.1 -3.1
Repeatability ±mg 0.001 0.001 0.003/0.004/0.005
Linearity ±mg 0.004 0.004 0.004
Corner load (test load [g])
1
) μg 4 (2 g) 4 (2 g) 5 (1 g)
Min. initial weight
2
) mg 2 – 4
Sensitivity drift between +10 to +30°C ±ppm/K 1 1 1
Typical stabilization time s 5 5 5
Typical measurement time s 8 8 8
External standard calibration value g 5 (E2) 5 (E2) 3 (E2)
(of at least accuracy class...)
Display result
(depending on the set filter level) 0.1 – 0.4 0.1 – 0.4 0.1 – 0.4
Weighing pan size mm 30 50/30 30
Weighing chamber height mm 70 15 70
Protection Protected against dust and water
Ultramicrobalances 0.0001mg
Model MSE2.7S MSE2.7S-F
Readability mg 0.0001 0.0001
Weighing capacity g 2.1 2.1
Tare range (subtractive) g –2.1 –2.1
Repeatability ±mg 0.00025 0.00025
Linearity ±mg 0.0009 0.0009
Corner load (test load [g])
1
) μg 0.5 (1 g) 0.5 (1 g)
Min. initial weight
2
) mg 1 –
Sensitivity drift between +10 to +30°C ±ppm/K 1 1
Typical stabilization time s 7
Typical measurement time s 10
External standard calibration value g 2 (E2) 2 (E2)
(min. accuracy class)
Display result
(depending on the set filter level) 0.1 – 0.4 0.1 – 0.4
Weighing pan size mm 20 50/20
1
)
Weighing chamber height mm 70 15
Protection Protected against dust and water
1
) = Standard pan
2
) = Typical min. initial weighing according to USP (United States Pharmacopeia), USP31-NF26
88 Cubis MSE Operating Instructions