Operating instructions

Cubis MSE Operating Instructions 71
Data Interfaces
Data Input
SBI Commands (Data Input Format)
The computer connected via the data port can send control commands to the
balance to control balance and application program functions.
These control commands may have different formats and contain up to 20
characters. Each of these characters must be sent based on the setup configuration
for data transmission.
Formats for Control Commands (Syntax)
Format 1: Esc ! CR LF
Format 2: Esc ! # _ CR LF
Esc: Escape
!: Command character
#: Digit
&: Parameter (number or letter)
_: Underline (ASCII: 95)
CR: Carriage return (optional)
LF: Line feed (optional)
Examples:
Format 1: Esc P
Format 2: Esc x1_