Operating instructions
Configuration (Operating Menu)
38 Cubis MSE Operating Instructions
Parameter Settings: Overview
k = Factory setting 3 = User-defined setting
Info about
Level 1 Level 2 Level 3 Level 4 Menu Level
1) Setup BAL.SCAL. AMBIENT conditions (adapt filter) V.Stable Very stable 1. 1. 1. 1
Balance k STABLE 1. 1. 1. 2
parameters UNSTABL 1. 1. 1. 3
V.UNSTBL. Very unstable 1. 1. 1. 4
App.Filt. Application filter k FINAL.RD. Final readout mode 1. 1. 2. 1
Filling Filling mode 1. 1. 2. 2
REDUC. Reduced 1. 1. 2. 3
OFF 1. 1. 2. 4
Stab.Rng. Stability range Max. ACC. Maximum accuracy 1. 1. 3. 1
V.ACC. 1. 1. 3. 2
k ACC. 1. 1. 3. 3
FAST 1. 1. 3. 4
V.FAST. 1. 1. 3. 5
Max.FAST
1
) 1. 1. 3. 6
ST.DEL. Delay No 1. 1. 4. 1
k Short 1. 1. 4. 2
AVERG. 1. 1. 4. 3
LONG 1. 1. 4. 4
Tare
1
) WIOSTB.W/o stability 1. 1. 5. 1
k WISTB.After stability 1. 1. 5. 2
ATSTAB. At stability 1. 1. 5. 3
Aut.zero Off 1. 1. 6. 1
Auto zero k on 1. 1. 6. 2
WT.UNIT For list of units, see 1. 1. 7. 1
Basic weight unit “Toggling Between Weight Units” to
1. 1. 7.24
DISP.DIG.
1
) k All 1. 1. 8. 1
Display accuracy lp.on.off (last digit after load change) 1. 1. 8. 2
DIV.1 1 interval 1. 1. 8. 6
Minus 1 Reduced by 1 digit 1. 1. 8. 7
Cal./ADJ. EXT.CAL. External calibration with standard weight
2
) 1. 1. 9. 1
Function of C key e.cal.usr. External calibr./adjustment with user-defined weight
2
) 1. 1. 9. 3
k Cal.Int. Internal calibration 1. 1. 9. 4
Int.lin. Internal linearization (on analytical balances only) 1. 1. 9. 5
Set.Prel. Setting the Preload 1. 1. 9. 8
Del.Prel. Clearing the Preload 1. 1. 9. 9
blocked C Blocked 1. 1. 9.10
Select 1. 1. 9.12
set.ext.w. Determine ext. calibration weight for e.cal.usr. 1. 1. 9.17
Cal.adj.ON k Sequence adjustment 1. 1.10. 1
CAL.ADJ. Adjustment as needed 1. 1.10. 2
Zero rng. Zero range 1 perc.ent 1. 1.11. 1
k 2 perc.ent 1. 1.11. 2
INIT.ZERO Zero at Power On k default Factory setting 1. 1.12. 1
2 perc.ent 1. 1.12. 2
On Tare Tare/zero at power on k ON 1. 1.13. 1
OFF 1. 1.13. 2
CYC.rate Output rate k normal 1. 1.14. 1
highvar. 1. 1.14. 2
slow 1. 1.14. 3
medium 1. 1.14. 4
fast 1. 1.14. 5
veryfast 1. 1.14. 6
maximum 1. 1.14. 7
ISOcal Autom. calibration/adjustment Off 1. 1.15. 1
Note 1. 1.15. 2
k On 1. 1.15. 3
To.w.linearization (for analytical balances only) 1. 1.15. 4
EXT.Cal. External calibration
1
) k Free 1. 1.16. 1
Locked (blocked) 1. 1.16. 2
cal.unit
1
) k Gram 1. 1.17. 1
for calibration weight Kilogr. Kilograms 1. 1.17. 2
USERDEF. (factory setting: pound) 1. 1.17. 4
GEN.SERV. men.Reset Menu reset Yes Restore fcty. settings 1. 9. 1. 1
General servicing (Factory Settings) k no Do not restore settings 1. 9. 1. 2
Standard
1
) 1. 9. 1. 3
Verifiable
1
) 1. 9. 1. 4
1
) Not available on balances verified for use in legal metrology
2
) Not available on balances for use in legal metrology of accuracy class K