Operating instructions
32 Cubis MSE Operating Instructions
Operation
Opening and closing the motorized draft shield (optional)
*
If your balance is equipped with a motorized draft shield, you can use the * key
to open and close the draft shield doors.
Models with the analytical draft shield:
The * key can be set, for example, so that it opens and closes the right and
top door (for right-handed people). The opening width (entirely or partially) of the
sliding panels can also be defined. You can configure these settings manually on
the draft shield itself (see next section “Learning Function" below).
Assigning Opening Functions to the * Key (Learning Function)
The balance must be turned on.
t Close all draft shield doors.
t Manually operate the door you want to assign to the * key: Push the right
door open as far as you want it to opened (entirely or partially).
t Press the * key.
y The door will close again. Now you have configured the * key so that, when
pressed, it opens or closes the right door.
t Proceed in the same way to configure the left door.
h
Using the same method, you can assign the upper door to this key. You can also
configure one key to move both doors by manually opening both doors during
setup before pressing the * key.
Models with a rotation draft shield (MSx6.6S/3.6P/2.7S):
t Set the opening width: manually select the desired opening angle
t 1) Open to the previously selected opening angle:
Press the * key
2) Close: Press the * key
100°
180°
260°
0°