Basic Weighing Function
Features
Tare the balance
Print weight value
Using a verified balance in legal metrology:
The Type-Approval Certificate for verified balances is only valid for non-automatic
weighing instruments.
For automatic operation with or without additional, integrated equipment, please
follow the applicable national regulations for the installation location.
§ Before use in legal metrology, the balance should be adjusted via the internal
motorized calibration weight at the installation location:
see the “Calibration and Adjustment" section in this chapter.
$
The temperature range (°C) listed on the ID label should not be exceeded during
operation.
Example:
MSX
K
+10/+30°C
+5°C/+40°C isoCAL
Preparation
§ Turn balance on: Press the A key
§ Level the balance if required
§ Tare the balance if necessary: Press the J key
$ If necessary, change the configuration settings: see “Configuration"
$ If desired, load the factory settings: see “Configuration"
Additional functions:
$ Turn balance off: Press the A key
Quick Guide: First Weighing
Step Press key Display/Printout
1. Turn balance on A 0.0 g
Self-test runs.
Then the balance performs an automatic initial tare.
2. Place container on weighing pan (in this example : 11.5 g). + 11.5 g
Close the draft shield (if available).
3. Tare the balance J 0.0 g
The balance is tared and the displayed value is zero.
4. If required, open the draft shield and carefully place the sample in
the container on the weighing pan - in this example : 132 g. + 132.0 g
Close the draft shield (if available). You can read the measured value
as soon as the weight value stops changing and the unit is displayed.
5. Print weight value P N + 132.0 g