Getting Started
Setting up a filter weighing pan for left-handed users:
t Remove the draft shield cover
t Unscrew the pin and move from the right to the left side
t Turn draft shield parts approx. 90 degrees to the left (loosen knurled screw)
Power Connection
t Check the voltage rating and plug design.
– Make sure that the voltage rating printed on the manufacturer’s ID label is
identical to that of your local line voltage.
– If the stated supply voltage or the plug design of the power cord does not
comply with your country’s standard, please inform the nearest Sartorius
representative or your dealer.
– The power connection must be made in accordance with the regulations
applicable in your country.
– In order to connect the equipment to the power supply (protection class 1), use
a suitable, correctly installed wall outlet with a protective grounding conductor
(PE) and a fuse of a maximum 16 A.
– If a main feeder from the ceiling is required or if installation of a CEE plug is
necessary, these must be carried out by a qualified electrician.
– Only use original Sartorius AC adapters.
16 Cubis MSE Operating Instructions