Service manual
99
The grounding terminal is identified by
a grounding (earthing) symbol.
The load cells are made of stainless steel and
designated as EX (see right side).
The spare part numbers are different.
There are platforms with 4 load cells (see left
side); see photo showing the type of one of
the load cells installed (see right side).
The junction box and the pin assignment of
the cables are identical.
The output cable routed to the indicator is
identified by blue shrink-on tubing.
There is a platform with 1 load cell (see left
side); see photo showing the type of load cell
installed (see right side)
The service (calibration and adjustment) data
are identical.
Warning!Warning!
Warning!Warning!
Warning!
Only an EX indicator may be connected toOnly an EX indicator may be connected to
Only an EX indicator may be connected toOnly an EX indicator may be connected to
Only an EX indicator may be connected to
the platform scale.the platform scale.
the platform scale.the platform scale.
the platform scale.