Service manual
98
CAPXS… Explosion-Protected Models
Warning! Only technicians with EX authorization and certification may service this
explosion-protected (EX) equipment. Afterwards, the scales must be resealed
using the adhesive EX control marks with the technician’s personal ID printed
on them.
Service Concept: On EX models,
only the complete subassemblies and not the components are
exchanged. After you have exchanged a subassembly, such as the PCB or
display unit, you may not perform any high-voltage tests because the scales
still have a potential field. A high-voltage test would destroy the electronic
components!
Important Note: For hazardous areas, weighing platforms are available in the stainless steel versions
only.
What Is Different in Comparison to Other EX Models?
Ex manufacturer’s type labels are on the equipment
AUT_9852a.jpg
AUT_9863a.jpg