Service manual
83
– When installing the new load cell, make sure any arrow (see the illustration on
the left) always points in the movement of the load cell.
– Tighten the load cells with the required torque.
Torque Values
– After replacing load cells, you need to check the overload stops and adjust
them if necessary (see page 77-80).
lastz_2.jpg
lastz_3.jpg
Load cell designation Torque Load cell designation Torque
GWT Type 011231/550 C3 130 Nm GWT Type 011290/500 C3 35 Nm
GWT Type 011232/1100 C3 130 Nm GWT Type 011293/7,5 C3 10 Nm
GWT Type 011233/1760 C3 130 Nm GWT Type 011294/15 C3 10 Nm
GWT Type 011241/7,5 C3 10 Nm GWT Type 011295/30 C3 10 Nm
GWT Type 011242/15 C3 10 Nm GWT Type 011296/50 C3 10 Nm
GWT Type 011243/30 C3 10 Nm GWT Type 011297/100 C3 10 Nm
GWT Type 011244/50 C3 10 Nm GWT Type 011298/200 C3 20 Nm
GWT Type 011245/100 C3 14 Nm GWT Type 011299/7,5 C3 10 Nm
GWT Type 011246/50 C3 14 Nm GWT Type 011300/15 C3 10 Nm
GWT Type 011247/100 C3 14 Nm GWT Type 011301/30 C3 10 Nm
GWT Type 011248/200 C3 14 Nm GWT Type 011302/50 C3 10 Nm
GWT Type 011249/100 C3 35 Nm GWT Type 011303/100 C3 10 Nm
GWT Type 011250/200 C3 35 Nm GWT Type 011304/200 C3 20 Nm
GWT Type 011251/50 C3 35 Nm GWT Type 011305/50 C3 35 Nm
GWT Type 011252/150 C3 35 Nm GWT Type 011306/100 C3 35 Nm
GWT Type 011253/250 C3 35 Nm GWT Type 011307/200 C3 35 Nm
GWT Type 011272/50 C3 35 Nm GWT Type 011308/500 C3 35 Nm
GWT Type 011273/100 C3 35 Nm GWT Type 011309/550 C3 130 Nm
GWT Type 011274/200 C3 35 Nm GWT Type 011310/1100 C3 130 Nm
GWT Type 011275/500 C3 35 Nm GWT Type 011311/1760 C3 130 Nm