Service manual
77
Setting the Overload Stops
1. Weighing Platforms with One Load Cell: Steel
Loosen the overload stops until the weight value displayed for the load on the
platform begins to decrease (stops touch the load plate support).
Then retract (tighten) the stops by 4 turn.
Platform nominal Test weight for Test weight for
capacity in kg middle stops in kg corner stops in kg
344
688
15 20 15
30 40 30
60 80 60
150 200 150
300 400 300