Adjusting the Weighing Platform
Adjusting the Off-center Load
If the weighing platform has only one load cell, the off-center load is not adjusted
in case of error; rather, the load cell must be replaced.
The off-center load is adjusted only in weighing platforms that have four load
cells.
Procedure (Example Illustrated on the Left)
Place the test weight (12.5 t) on the weighing platform and write down the
weight values measured.
The load cell with the lowest weight value (12450 ) is the reference load cell.
This load cell does not require adjustment.
Calculate the difference between the weight value registered by this load cell
and those registered by the other load cells.
Plot the point of intersection of the calculated off-center load difference
(x-axis) and the test weight (see page 75).
The adjustment resistance (total) and the best resistor combination can be read
from the y-axis (see the diagram on page 75).
In the example given here, load cell 3 has an off-center loading error of 80 kg,
which yields an adjustment resistance of 1.5 ohm + 0.82 ohm.
Depending on the resistor combinations of each individual load cell, the
required resistors are activated by disconnecting the corresponding “0-ohm
resistor” (see page 76).
Note: The resistance values detected are based on 350-ohm load cells.
For 700-ohm load cells, double the resistance values.
Eckl_2_e.eps