Service manual

72
1. Disconnect the platform measuring voltage lines from the terminal strip
(pins 3 and 4) in the A/D converter. Then short circuit these two lines and
connect them to pin 7 (GND) of the terminal connector. If the A/D converter is
intact, the weight readout should be stable.
Disconnect the cable from the platform.
2. Connect the strain-gauge simulator and turn the switch to simulate different
input signals. The weight readout should be stable.
Replacing the Digital PCB
Note: If communcation problems occur during communication with connected devices
(such as a printer), check the cable connections and the data transfer parameters
first; if the error source is not found, start the internal test program.
Turn off the indicator.
Press and hold the k key and turn the indicator on again. The test program
starts.
In case of defect, replace the digital PCB.
Some Combics indicators are equipped with D-Sub 25 female connectors, and
some with cable glands; make sure you order the correct PCB for your indicator
(see the illustration on the left).
After replacing the digital PCB, use the PPLoader program to load the
application software, if necessary.
Afterwards, enter the model designation and the Serial number of the complete
Combics scale (see pages 63 and 65).