Service manual

60
Press ) to store the A/D converter configuration data.
Busy” is displayed briefly; afterwards, the Combics is re-started.
After the self-test of the display, the scale is ready to operate.
Menu Item 1-19: Entering the Specifications of the Load Cell(s)
Activate the service mode (see page 15).
Select weighing platform »WP1«.
Press the ) key to open the numeric menu.
Press the ) key, then seveal times, k ) to open menu line 1-19,
“Calibration without weights”; then select item 1-19-1, “Nominal capacity”.
Press ) again to activate menu item 1-19-1.
Enter the nominal capacity of the load cell(s) (e.g., 30.00 kg).
The nominal capacity has to match the maximum load set under
menu item 11-4-4.