Service manual

57
Press ) to confirm your selection.
In the example shown here, menu item 11-3-1 (Single-range scale) has been
selected.
If desired, press k to change the weighing range structure:
11-3-2: Multiple-range scale; 11-2-3: Multi-interval scale.
Press ( to return to menu line 11-3.
Note: Entering the scale interval, range limits (multiple-range or multi-interval scales
only) and maximum capacity:
The menu line corresponding to the selected weighing range structure
is displayed, i.e.,
– for single-range scale (menu item 11-3-1): in relation to menu item 11-4;
– for multiple-range scale (menu item 11-3-2): in relation to menu item 11-5;
– for multi-interval scale (menu item 11-3-3): in relation to menu item 11-6.
In each case, the menu lines for the other two weighing range structures are not
shown.
When you activate menu code 11-3-1 (Single-range scale) (by pressing the
key), menu line 11-4 is opened automatically. In this case, menu lines 11-5 and
11-6 are not shown.
Open menu item 11-4-1 to enter the scale interval d.
Press ) to select the menu item.
display_serv_1131.eps.
menu_1141.eps.
display_serv_114.eps.