Service manual

41
Enter the acceleration of gravity at the place of installation (in ms
-2
/ or the
reference value: 9.811590). Press ) to confirm your input.
Activate menu item 1-20-4 to store new data (key sequence: ) k).
The display shows “busy”, indicating that the parameters entered are being stored.
Restart the scale: Turn the indicator off and then on again (key sequence: e
e, or press and hold the ) key (> 2 sec).
After the self-test of the display, the scale is ready to operate.
Note: Once the precise geographical data or acceleration of gravity has been entered,
the tolerance range is restricted (see page 23).
A new opening of this menu item will show 0.00... in the display only.
After data has been entered as described above, the span must
be adjusted.
Example
Settings (changes in the factory settings required for this example);
Menu item 1-9-1: External calibration/adjustment with default weight;
Menu item 1-10-1: Calibration/adjustment in one operation;
Menu item 1-16-1: External calibration/adjustment/linearization accessible;
display_busy.eps
display.eps