Service manual

34
Adjustment without Weights
Menu Item 1-19: Entering the Specifications of the Load Cell(s)
Activate the service mode (see page 15).
Select weighing platform »WP1«.
Press the ) key to open the numeric menu.
Press the ) key, then, several times, k repeatedly and then ) again to
open menu item 1-19, “Calibration without weights”; then select item 1-19-1,
“Nominal capacity”.
Press ) again to activate menu item 1-19-1.
Enter the nominal capacity of the load cell(s) (e.g., 30.00 kg).
The nominal capacity has to match the maximum load set under menu item
11-4-4.
Note: If a weighing platform has more than one load cell, multiply the nominal capacity
accordingly. Example: The weighing platform consists of 4 load cells, each of them
has a capacity of 30 kg.
In this case, the nominal capacity (= maximum load) equals 4 x 30 kg, or 120 kg.
display_wps1_serv.eps
display_serv_01.eps
display_serv_01191.eps
display_mxl30.eps