Service manual

27
Press ), then, several times, k ) to select menu item 1-9-3, “External
calibration/adjustment with user-defined weights” and confirm by pressing )
briefly.
Press ( to jump to the first columns of the numeric menu.
Press ) to store the selected menu item.
Turn the Combics indicator off and then on again to exit the Service mode.
After taring or zeroing the scale, press and hold the ) key until Cal is
displayed.
If menu item 8-12-2 is active, the geographical data is displayed before the
calibration weight is prompted. Confirm each of the values shown by pressing
the ) key to continue with the calibration routine.
After approx. 2 seconds the calibration weight (user-defined weight) required
is shown on the display. Select menu item “1-18-1” to enter the user-defined
weight.
Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last span adjustment is displayed (calibration).
Press the ( key to stop the calibration/adjustment routine.
Press the ) key to start the adjustment.
cal.eps
10kg.eps
display_serv_01.eps
10kgw.eps
menu_193.eps