Service manual

26
cal.eps
5kg.eps
display_serv_01.eps
5.0kgw.eps
display_wps1_serv.eps
display_serv_01.eps
After taring or zeroing the scale, press and hold the ) key until Cal is
displayed.
If menu item 8-12-2 is active, the geographical data is displayed before the
calibration weight is prompted. Confirm each of the values shown by pressing
the ) key to continue with the calibration routine.
After approx. 2 seconds the calibration weight required is shown on the display.
Place the prompted weight on the scale. After a brief pause, the difference
between the current value and the last is displayed (calibration).
Press the ( key to stop the calibration/adjustment routine.
Press the ) key to start the adjustment.
1-9-3 External Calibration/Adjustment with User-defined Weights
Select weighing platform »WP1«.
Press the ) key to open the numeric menu.