Menu Item 1-18: Entering Calibration/Linearization Weights
1-18-1: External User-defined Calibration Weight (Service Mode not Required)
Activate the service mode (see page 16).
Select weighing platform “WP1”.
Press the ) key to open the numeric menu.
Select menu item 1-18-1 (external user-defined weight) by pressing k )
repeatedly.
Activate input mode for the user-defined weight (press ) briefly).
The first digit flashes.
Enter the user-defined weight (in this example: 10,000 kg) (key sequence:
) k ) p ) ) ) k )). The last number entered flashes.
Press the ) key again to store the user-defined weight; the menu jumps back
to 1-18-1.