Service manual

17
Working in the Service Mode
Note: When performing maintenance on an existing system, the service mode must be activated to perform
linearization, to set or clear the preload, and to enter the date of service.
Additional Menus in Service Mode
In the Text Menu (“Setup”) Under the “Date” (DATE) and “Password” (CodE) menu items:
Date of service S-DATE
Memory number MEM-NO
Indicator serial number SER-NO
Model designation ModEL
In the numeric menu: The Setup menu for WP1 (WP-1) and the COM1 (COM1) and UniCOM (UNICOM) ports is
extended by the following items for configuring weighing platforms WP1 (WP-1) and WP2 (WP-2):
1-9 Calibration/adjustment functions:
1-9-1: Ext. calibration/adjustment with default weight
(service mode not required)
1-9-3: Ext. calibration/adjustment with user-defined weight
(service mode not required)
1-9-5: Internal linearization (only for WP2 on COM1 and UniCOM ports)
1-9-7: External linearization with user-defined weights
1-9-8: Set preload
1-9-9: Clear preload
1-9-10: Key blocked (service mode not required)
IF „calibration“ first, after that „span adjustment“ is configured in the menu (1.10.2), carry on to put necessary
weights onto the load plate even if the difference to the last span adjustment (calibration function) is already
displayed. When all weights are loaded, press the „tare“ button to start the adjustment function.