Operating instructions
4
Counting
Z
Preparation
§ Select Setup: Press the M key
§ Select Application Parameters:
Press the
O soft key
§ Select the Counting application:
Press the
O soft key
Counting
Minimum load for autotaring
1 digit
2 digits
5 digits
o 10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Minimum load for initialization
o 1 digit
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
Autotare first weight
o Off
On
Autostart app. at power on
On
o Off
Accuracy: avg. piece wt. calculation
o Display accuracy
Display accuracy + 1 decimal place
Display accuracy + 2 decimal places
With internal resolution
Parameter for saving weight
o Standard stability
Increased stability
Average piece weight updating
Off
o Automatic
Scale for reference weight
o Do not change
WP 1
WP 2
Factory settings
Yes
oNo
o = factory setting
§ Save settings and exit Setup:
Press the M key or the
oo soft key
Parameter for Saving Weight Values
The weight on the platform is saved
as a reference value when the platform
has stabilized. “Stability” is defined as
the point at which the fluctuation of
a measured value lies within a defined
tolerance range. The narrower the
tolerance range, the more stable the
platform is at “stability”.
In Setup, under:
Application:
Counting: Save wt.,
you can define whether the value is
saved when “standard stability” is
reached, or only at “increased stability”
(narrower tolerance range). If you
select
increased stabili-
ty, the value saved for average piece
weight will be more accurate and the
results more reproducible, but the
response time of the weighing platform
might be longer.
Accuracy of Average Piece Weight
Calculation
The resolution applied for calculating
the average piece weight is defined in
Setup under:
Application:
Counting: APW calcul.
The resolution for calculating the
average piece weight is increased if
“+1 decimal place”, “+2 decimal places”
or “Internal resolution” is selected.
With the “+1 decimal place” setting,
the net value is determined to one
additional decimal place (i.e., display
accuracy x 10); “+2 decimal places”
increases display accuracy + 100, and
so on up to the maximum resolution
available.
Minimum Load
The minimum load required for initial-
ization of the weighing platform is con-
figured in Setup under:
Application:
Counting: MinL.Init
Once the limit is exceeded by the load,
initialization can begin. If the load on
platform is too light, the following will
occur when you try to save the refer-
ence value:
– The error code
inf 29 is displayed
– A warning signal is emitted (double-
beep)
– The weighing platform is not initialized
– The preset reference sample quantity is
saved
The minimum load required for auto-
matic taring of the container weight on
the platform (first weight) is configured
in Setup under:
Application:
Counting:
MinL.Autotar
You can choose from the following
10 levels for this setting:
1 digit (no minimum load)
2 digits
5 digits
10 digits
20 digits
50 digits
100 digits
200 digits
500 digits
1000 digits
The “digits” here refer to the scale
intervals in the connected weighing
platform. If the interval of the con-
nected platform is 1 g, for example,
and 1000 digits are required, you must
place at least 1000 g (=1000 intervals
=1000 digits) on the weighing platform
for initialization.